Original Research
December 01, 2013
JTCB. 2013 Dec 1; 11 (3) : 099-104

Comparison of Primus Ultra2 Analyzer and Tosoh HLC-723 G8 HPLC Systems in Thalassemia Screening

1 Antalya Eğitim ve Araştırma Hastanesi, Tıbbi Biyokimya Laboratuvarı, Antalya
2 Antalya Atatürk Devlet Hastanesi, Tıbbi Biyokimya Laboratuvarı, Antalya

Abstract

Objective: HPLC (High Performance Liquid Chromatography) is the most commonly used method to differentiate the subtypes of hemoglobin in screening for hemoglobinopathy. Today, different models and trademarks of HPLC equipment have been used in screening for hemoglobinopathy. However, the agreement of the results of the analyses performed by these types of equipment are unknown. In this study, the aim was to investigate the agreement between the results of the analyses of hemoglobinopathy screening performed in two sets of HPLC equipment with the Primus Ultra2 analyzer and the analyzer Tosoh HLC 723 G8 trademarks.

Materials and Methods: A total of 40 individuals, 22 males and 18 females, presented to the laboratory between March 15 and April 15, 2013 for thalassemia screening were included in this prospective study. Statistical correlations of hemoglobin A2 (HbA2) values, analyzed in each analyzer as a single analysis from each sample were evaluated using student t test, two-way random intraclass correlation coefficient (ICC), concordance correlation coefficient, correlation coefficient (r) and regression analysis.

Results: As a result of the analysis, 6 of the samples (33.7%) were detected as -thalassemia carrier, 10 (36.2%) were diagnosed as anemic and 24 (30%) were detected to be normal. HbA2 values were not significantly different between the two HPLC equipments (t= -1.269 p= 0.208). Correlation coefficients for HbA2 values were found to be as r= 0.987, ICC=0.97 and CCC= 0.93. Regression analysis for HbA2 values were found to be as Y= -0.0605+1.1613X.

Conclusion: According to the regression, correlation and other statistics analysis, agreement between the results of HbA2 values measured in the two analyzers was found to be sufficient.

Citation

Ellidağ, Hamit Yaşar, Esin Eren, Serpil Aktaş Tosun, and Necat Yılmaz. 2013. “Comparison of Primus Ultra2 Analyzer and Tosoh HLC-723 G8 HPLC Systems in Thalassemia Screening”. Journal of Turkish Clinical Biochemistry 11 (3):099-104. https://turkclinbiochem.org/index.php/pub/article/view/190.

Publish with JTCB

Diamond Open Access

Fully Diamond OA with no APCs and no submission fees. Your research is freely available to readers worldwide from the day it publishes.

30-Day First Decision

Double-blind peer review from an active editorial team returns a first decision in about 30 days, keeping your manuscript moving.

Global Platform

Backed by the Turkish Society of Clinical Biochemistry, JTCB welcomes submissions from researchers worldwide and reaches an international laboratory audience.

Focused Laboratory Scope

Method validation, analyzer comparisons, reference intervals, quality indicators, preanalytical studies, and instructive clinical case reports.

Credible and Discoverable

All manuscripts publish in English and follow ICMJE and COPE guidelines, with Crossref DOIs and TR Dizin indexing ensuring global discoverability.