The Effects of Blood Collection Tubes on Folate Measurements

1 Çorum Eğitim ve Araştırma Hastanesi, Biyokimya Laboratuvarı, Çorum, Türkiye
2 Hitit Üniversitesi Tıp Fakültesi, Tıbbi Biyokimya Anabilim Dalı, Çorum, Türkiye

Abstract

Objective: Blood collection tubes with a clot activator are widely used in clinical laboratories. Unexpectedly high results in folate assays obtained on Siemens Advia Centaur XP analyzer triggered a need for an investigation of the effects of blood collection tubes on folate measurements.

Material and Methods: We investigated the potential assay interference of three types of plastic tubes with a clot activator; Vacutest REF11030 (Group A); Becton Dickinson (BD) Vacutainer CAT REF367896 (Group B);and BD Vacutainer SSTTM II Advance REF367953 (with gel) (Group C) tubes. Blood samples from apparently healthy 13 volunteers were drawn into these different tube types and serum folate concentrations were analyzed by Siemens Advia Centaur XP, Roche E-170, and Beckmann UniCel DxI800 immunoassay analyzers. The results were compared and p values <0.05 were considered as statistically significant.

Results: Folate values measured one Siemens Advia Centaur XP, were significantly higher in Group A (13.44 ± 4.56 ng/mL) than Group B (8.82 ± 3.33 ng/mL) or Group C (8.58 ± 2.83 ng/mL) tubes (p<0.001). No difference was found between the tube groups regarding Roche E-170 and Beckmann UniCel DXI800 analyzers (p>0.05). When serum folate levels drawn into Grup A tubes were compared with different analyzers, results obtained from Siemens Advia Centaur XP (13.44 ± 4.56 ng/mL) were higher than results obtained from Roche E-170 (9.15 ± 2.25 ng/mL) or Beckmann Unicell DXI800 (8.70 ± 3.00 ng/mL) analyzers (p<0.001).

Conclusion: Siemens folate reagent was significantly influenced by the contents of the Vacutest tube with a clot activator resulting in high results. Assay interferences born from blood collection tubes represent challenges to clinical laboratories because they are not detected by the usual quality control programs. Therefore, when starting to use a new blood collection tube, assay interferences should be tested, especially for immunoassay systems.

Citation

Erdemli, Hacı Kemal, Ramazan Kocabaş, and Sumeyya Akyol. 2016. “The Effects of Blood Collection Tubes on Folate Measurements”. Journal of Turkish Clinical Biochemistry 14 (1):001-005. https://turkclinbiochem.org/index.php/pub/article/view/237.

Publish with JTCB

Diamond Open Access

Fully Diamond OA with no APCs and no submission fees. Your research is freely available to readers worldwide from the day it publishes.

30-Day First Decision

Double-blind peer review from an active editorial team returns a first decision in about 30 days, keeping your manuscript moving.

Global Platform

Backed by the Turkish Society of Clinical Biochemistry, JTCB welcomes submissions from researchers worldwide and reaches an international laboratory audience.

Focused Laboratory Scope

Method validation, analyzer comparisons, reference intervals, quality indicators, preanalytical studies, and instructive clinical case reports.

Credible and Discoverable

All manuscripts publish in English and follow ICMJE and COPE guidelines, with Crossref DOIs and TR Dizin indexing ensuring global discoverability.