Journal of Turkish Clinical Biochemistry

The official publication of Turkish Society of Clinical Biochemistry

Diamond OA
No APCs
First Decision
30 Days
Indexed In
TR Dizin
Original Research
April 01, 2016
JTCB. 2016 Apr 1; 14 (1) : 051-057

Immunoturbidimetric and Electrochemiluminescence Methods for Ferritin Measurement

1 Tepecik Eğitim ve Araştırma Hastanesi, Tıbbi Biyokimya, İzmir, Türkiye
2 Antalya Eğitim ve Araştırma Hastanesi, Tıbbi Biyokimya, Antalya, Türkiye
3 Dr. Suat Seren Göğüs Hastalıkları ve Göğüs Cerrahisi, Tıbbi Biyokimya, İzmir, Türkiye

Abstract

Purpose: The measurement of serum ferritin is used as a reliable and noninvasive method for the monitoring of iron stores. There are many different methods for the ferritin measurement. However, the immunoturbidimetric method that having the less cost and possibility of the simultaneously ferritin measurement with other parameters is more useful than electrochemiluminescent method. Thus, we aimed to compare the results of immunoturbidimetric and electrochemiluminescent method.

Material and Methods: The 109 blood samples were analyzed by immunoturbidimetric and electrochemiluminescent method. The comparison study was performed according to the Clinical Laboratory Standards Institute EP9-A3 protocol. The bias between two methods was evaluated according to acceptable systematic error limit. Diagnostic agreement according to the reference value was analyzed by the Kappa test.

Results: Although a strong positive correlation (r=0.999; p<0.001) was detected between the results of immunoturbidimetric and electrochemiluminescent method, it was found a significant difference between the results of two methods (p<0.001). This difference was within the acceptable limits according to the bias obtained from various sources. The agreement of diagnostic according to reference values was good (Kappa value, 0.62; p<0.001) in women; fair (Kappa value, 0.28; p<0.001) in men.

Conclusions: Although there was analytical agreement between immunoturbidimetric and electrochemiluminescence method, it should be considered that the difference may occur at high ferritin levels. It can increase the degree of diagnostic agreement determining the appropriate reference intervals for laboratory's population instead of reference intervals provided by the manufacturer. So, immunoturbidimetric method which having practical advantages may be preferred instead of electrochemiluminescence method.

Citation

İnce, Fatma Demet, Hamit Yaşar Ellidağ, Merve Zeytinli, Ramazan Avcı, Defne Özbek, and Elif Merve Arı. 2016. “Immunoturbidimetric and Electrochemiluminescence Methods for Ferritin Measurement”. Journal of Turkish Clinical Biochemistry 14 (1):051-057. https://turkclinbiochem.org/index.php/pub/article/view/245.

Publish with JTCB

Diamond Open Access

Fully Diamond OA with no APCs and no submission fees. Your research is freely available to readers worldwide from the day it publishes.

30-Day First Decision

Double-blind peer review from an active editorial team returns a first decision in about 30 days, keeping your manuscript moving.

Global Platform

Backed by the Turkish Society of Clinical Biochemistry, JTCB welcomes submissions from researchers worldwide and reaches an international laboratory audience.

Focused Laboratory Scope

Method validation, analyzer comparisons, reference intervals, quality indicators, preanalytical studies, and instructive clinical case reports.

Credible and Discoverable

All manuscripts publish in English and follow ICMJE and COPE guidelines, with Crossref DOIs and TR Dizin indexing ensuring global discoverability.