Original Research
December 01, 2016
JTCB. 2016 Dec 1; 14 (3) : 181-188

Analytical Performance of Immuno Assay Method of 25 Hydroxy Vitamin D

1 SB 1 nolu Halk Sağlığı Laboratuvarı, Tıbbi Biyokimya, İstanbul, Türkiye
2 Haydarpaşa Numune Eğitim ve Araştırma Hastanesi, Tıbbi Biyokimya, İstanbul, Türkiye
3 Mağusa Tıp Merkezi Hastanesi, Tıbbi Biyokimya, Magusa, Kuzey Kıbrıs TC
4 Adana Halk Sağlığı Laboratuvarı, Tıbbi Biyokimya, Adana, Türkiye
5 Giresun Üniversitesi Tıp Fakültesi, Tıbbi Biyokimya, Giresun, Türkiye

Abstract

Aim: Method evaluation studies should be carried out for every new method planned to be used in laboratories. The aim of this study is to determine the analytical performance of the immunoassay method used for determining of plasma 25-hydroxy vitamin D levels and also to compare to high performance liquid chromatography (HPLC) method.

Materials and Methods: 113 plasma samples were used to carry out repeatability, linearity and recovery studies for 25-hydroxy vitamin D measurement with the Abbott Architect chemiluminescence immunoassay (CLIA) method. HPLC was accepted as reference method. Method comparison study was carried out.

Results: The coefficient of variations for the Architect 25-hydroxy vitamin D method were 4.07% and 2.42 % within-day; 4.77% and 3.93% between-day for low and intermediate plasma levels respectively. The results obtained from the recovery studies were 92.5% and 105.9% for low levels; and 108.3% and 105.5% for intermediate levels. As a result of the method comparison study, Architect method was found to have a significant negative bias [As mean, 9 ng/mL lower, ±2SD interval: (-34.1) – (16.1) ng/mL] and it was also observed that there was a constant and proportional difference between that two methods (HPLC= -3.2657 + 1.7388 CLIA; rho= 0.656).

Conclusion: Architect 25-hydroxy vitamin D method has met the precision, linearity and recovary targets but not the bias targets. For the accurate diagnosis and treatment follow-up of Vitamin D deficiency, it is obvious that to carry out analytical performance studies and to standardize of the measurement methods are essential. As the use of standard reference materials developed for the standardization of vitamin methods becomes widespread, the inter-method variability will also decrease. In addition, participation in laboratories' external quality assessment programs and their assessment of results will provide significant contributions to the use of laboratory findings for patient benefit.

Citation

Kulan, Pelin, Ahmet Rıza Uras, Sedef Delibaş, Mustafa Durmuşcan, and Sembol Yıldırmak. 2016. “Analytical Performance of Immuno Assay Method of 25 Hydroxy Vitamin D”. Journal of Turkish Clinical Biochemistry 14 (3):181-88. https://turkclinbiochem.org/index.php/pub/article/view/261.

Publish with JTCB

Diamond Open Access

Fully Diamond OA with no APCs and no submission fees. Your research is freely available to readers worldwide from the day it publishes.

30-Day First Decision

Double-blind peer review from an active editorial team returns a first decision in about 30 days, keeping your manuscript moving.

Global Platform

Backed by the Turkish Society of Clinical Biochemistry, JTCB welcomes submissions from researchers worldwide and reaches an international laboratory audience.

Focused Laboratory Scope

Method validation, analyzer comparisons, reference intervals, quality indicators, preanalytical studies, and instructive clinical case reports.

Credible and Discoverable

All manuscripts publish in English and follow ICMJE and COPE guidelines, with Crossref DOIs and TR Dizin indexing ensuring global discoverability.