The Comparison of Different Automation Systems with Reference Method for Erytrocyte Sedimentation Rate Measurement

1 Başkent Üniversitesi Alanya Araştırma ve Uygulama Merkezi Biyokimya Laboratuvarı, Ankara

Abstract

Purpose: In recent decades, several new techniques for measuring erytrocyte sedimentation rate have been developed in clinical laboratories. The aim of this study to compare three different automated sedimentation analyzers (Micro TEST-1, Sedisystem, Sed Rate Screener) with the manual Westergren reference method.

Materials and Methods: Patient blood samples were collected in K3EDTA vacutainers and in sodium citrate Vacu-Tec tubes and assayed within 4 hours of veinpuncture in three different sedimentation analyzer and manual Westergren method (n=40, 2-120 mm/h).

Results: The assays results of the automated sedimentation analyzers were statistically different from the manual Westergren method except Sedisystem (Sedisystem p=0.558, Micro TEST-1 p=0.034, Sed Rate Screener p=0.00 0). There were significantly high correlations between the results of sedimentation analyzers and the manuel Westergren method (Micro TEST-1 r= 0.712, Sedisystem r= 0.913, Sed Rate Screener r= 0.934, p values=0.000). The regression statistics obtained from the method comparison study were: Micro TEST-1; y = 0.80x + 0.56, Sedisystem; y = 1.10 x - 1.35, and Sed Rate Screener; y = 0.77x - 0.17.

Conclusion: The new fully automated systems for ESR measurement are safe, precise, accurate, and easy to use in clinical laboratories

Citation

Akalın, Nalan, Serap Arıkan, Nagihan Türkeli, and Turan Serter. 2004. “The Comparison of Different Automation Systems With Reference Method for Erytrocyte Sedimentation Rate Measurement”. Journal of Turkish Clinical Biochemistry 2 (2):077-082. https://turkclinbiochem.org/index.php/pub/article/view/34.

Publish with JTCB

Diamond Open Access

Fully Diamond OA with no APCs and no submission fees. Your research is freely available to readers worldwide from the day it publishes.

30-Day First Decision

Double-blind peer review from an active editorial team returns a first decision in about 30 days, keeping your manuscript moving.

Global Platform

Backed by the Turkish Society of Clinical Biochemistry, JTCB welcomes submissions from researchers worldwide and reaches an international laboratory audience.

Focused Laboratory Scope

Method validation, analyzer comparisons, reference intervals, quality indicators, preanalytical studies, and instructive clinical case reports.

Credible and Discoverable

All manuscripts publish in English and follow ICMJE and COPE guidelines, with Crossref DOIs and TR Dizin indexing ensuring global discoverability.