Vol. 19 No. 1 (2021), 050-055. Original Research. Published April 01, 2021.

Comparison of the Lifotronic H9 and Primus Ultra2 HPLC systems for the detection of HbA2

Mehmet Erten, Yeter Çakmak
  1. Halk Sağlığı Laboratuvarı, Tıbbi Biyokimya, Malatya, Türkiye

Corresponding author: Mehmet Erten, author@tkbdergisi.org

Abstract

Aim: Beta thalassemias are common autosomal recessive inherited hemoglobinopathy worldwide. Hemoglobin A2 levels are the most important parameter used to determine beta-thalassemia carriers. One of the most commonly used methods for HbA2 measurement is high-performance liquid chromatography (HPLC). We aimed to compare the HbA2 levels of Lifotronic H9 and Primus Ultra2 HPLC systems.

Material and Methods: A total of 51 individuals, 19 females and 32 males were included in our study. The compatibility of the results was demonstrated by the Bland-Altman plot and Passing-Bablok regression analysis.

Results: The geometric mean ratio of values was 0.95 (95% limits of agreement 0.6 to 1.49) in the Bland-Altman plot. 95% confidence intervals were observed as -1.03 - 1.30 (intercept) and 0.55-1.33 (slope) in Passing-Bablok regression analysis.

Conclusion: We reported first time measuring compatibility of Lifotronic H9 HPLC system in terms of HbA2 levels by using Primus Ultra2.

Keywords

How to Cite

Erten, Mehmet, and Yeter Çakmak. 2021. “Comparison of the Lifotronic H9 and Primus Ultra2 HPLC Systems for the Detection of HbA2”. Journal of Turkish Clinical Biochemistry 19 (1):050-055. https://turkclinbiochem.org/index.php/pub/article/view/385.

Publish with JTCB

Diamond Open Access

Fully Diamond OA with no APCs and no submission fees. Your research is freely available to readers worldwide from the day it publishes.

30-Day First Decision

Double-blind peer review from an active editorial team returns a first decision in about 30 days, keeping your manuscript moving.

Global Platform

Backed by the Turkish Society of Clinical Biochemistry, JTCB welcomes submissions from researchers worldwide and reaches an international laboratory audience.

Focused Laboratory Scope

Method validation, analyzer comparisons, reference intervals, quality indicators, preanalytical studies, and instructive clinical case reports.

Credible and Discoverable

All manuscripts publish in English and follow ICMJE and COPE guidelines, with Crossref DOIs and TR Dizin indexing ensuring global discoverability.