Journal of Turkish Clinical Biochemistry

The official publication of Turkish Society of Clinical Biochemistry

Diamond OA
No APCs
First Decision
30 Days
Indexed In
TR Dizin
Original Research
August 01, 2005
JTCB. 2005 Aug 1; 3 (2) : 045-050

The Performance Evaluation of Two Different Homocysteine Measurement Methods Used in Clinical Laboratories

1 Akdeniz Üniversitesi Tıp Fakültesi Merkez Laboratuvarı Biyokimya ve Klinik Biyokimya Ünitesi, Antalya
2 Akdeniz Üniversitesi Tıp Fakültesi Merkez Laboratuvarı Mikrobiyoloji ve Klinik Mikrobiyoloji Ünitesi, Antalya

Abstract

Objective: It has been proposed that the homocysteine (Hcy) produced during the metabolism of methionine plays a role in etiopathogenesis of several diseases. The interest in measurement of plasma Hcy levels has increased recently upon the finding that Hcy is an independent risk factor for cerebral, vascular and coronary artery diseases.

Materials and Methods: In the present study, we measured plasma Hcy levels with two different systems; Chromsystems kits in HP1100 and Drew DS30 and investigated the analytical performances of these two systems in Hcy measurements in clinical laboratories.

Results: Reproducibilities for Drew DS30 system and HPLC with Chromsystems kits were found to be 2.1 and 1.6%, respectively (p=0.043). Inter-assay variability for three different Hyc concentrations revealed standard deviations of 4.9 and 4.0% for low (5 µmol/L), 5.2 and 4.2% for moderate (12.5 µmol/L) and 5.8 and 4.2% for high (24.8 µmol/L) concentrations for Drew DS30 and HPLC with Chromsystems kits, respectively. Mean recovery was 101.8 ± 2.3% for Drew DS30 and 98.9 ± 1.4% for HPLC. The lowest measurable Hcy concentrations for Drew DS30 and HPLC systems were recorded as 2.8 and 1.6 µmol/L, respectively. There was a good correlation between the Hyc levels measured with both systems (r= 0.948). Limits of linearity were found to be 100 and 200 µmol/L for Drew DS30 system and HPLC with Chromsystems kits.

Conclusion: We demonstrated that the use of Chromsystems HPLC for homocysteine measurement might be more useful both in groups with predicted hyperhomocysteinemia and in laboratories with high number of patients where there are limitations in sample amounts, and working hours. On the other hand, in laboratories without the limitations mentioned above, DS30 system might be used with a similar performance to Chromsystems kits.

Citation

Özdem, Sebahat, S. Halide Akbaş, and Meral Gültekin. 2005. “The Performance Evaluation of Two Different Homocysteine Measurement Methods Used in Clinical Laboratories”. Journal of Turkish Clinical Biochemistry 3 (2):045-050. https://turkclinbiochem.org/index.php/pub/article/view/49.

Publish with JTCB

Diamond Open Access

Fully Diamond OA with no APCs and no submission fees. Your research is freely available to readers worldwide from the day it publishes.

30-Day First Decision

Double-blind peer review from an active editorial team returns a first decision in about 30 days, keeping your manuscript moving.

Global Platform

Backed by the Turkish Society of Clinical Biochemistry, JTCB welcomes submissions from researchers worldwide and reaches an international laboratory audience.

Focused Laboratory Scope

Method validation, analyzer comparisons, reference intervals, quality indicators, preanalytical studies, and instructive clinical case reports.

Credible and Discoverable

All manuscripts publish in English and follow ICMJE and COPE guidelines, with Crossref DOIs and TR Dizin indexing ensuring global discoverability.