Stability Studies of Urine Strips and Urine Samples

1 Atatürk Eğitim ve Araştırma Hastanesi, II. Klinik Biyokimya Laboratuvarı, İzmir

Abstract

Objectives: Urine specimen determinations sometimes may not be done immediately and may be studied by using urine strips which are ke pt in open air during routine work due to the conditions of study. In this study, the performance characteristics of urine strips and the changes in urine samples which were kept in open air were evaluated.

Materials and Methods: In the first step of study, 70 urine samples were analysed at 0, 2, 4, 6 th hours for the determination of changes in the samples.In the second step of study, urine strips kept in open air for 3 to 10 days were used. Urine specimens were analysed by both strips which were kept in original boxes and strips which were kept in open air.

Results: 100% false positive results were found for nitrite in urine strips kept in open air at 10th days.23% decrease in protein, 20% decrease in glucose and 33% negativity in keton measurements were observed in urine samples which were kept in open air for 6 hours. 10% decrease and 21% increase in density analysis, 23% decrease and 20% increase in pH analysis were found. Positivities were determined in nitrite analysis and bacteriuria (31%, 17% respectively) at 4 th hour in urine samples. There were 61% decrease in positivity of leukocyte at 6 th hour in specimens which were evaluated by both microscopy and strips. In the specimens which were evaluated by microscopy 75% decrease in positivity of erythrocyte was observed; however there were no changes in strip analysis.

Conclusion: In conclusion, according to these findings, it's very important to performe urine analysis in fresh urine as much as possible to achieve more reliable results. However, it was observed that accuracy in biochemistry analysis has not been affected by using urine strips stored in open air in short time intervals during intensive routine urine studies.

Citation

Topaç, Hatice, Aysel Hür, Ayşenur Atay, and Mehmet Köseoğlu. 2005. “Stability Studies of Urine Strips and Urine Samples”. Journal of Turkish Clinical Biochemistry 3 (2):067-072. https://turkclinbiochem.org/index.php/pub/article/view/52.

Publish with JTCB

Diamond Open Access

Fully Diamond OA with no APCs and no submission fees. Your research is freely available to readers worldwide from the day it publishes.

30-Day First Decision

Double-blind peer review from an active editorial team returns a first decision in about 30 days, keeping your manuscript moving.

Global Platform

Backed by the Turkish Society of Clinical Biochemistry, JTCB welcomes submissions from researchers worldwide and reaches an international laboratory audience.

Focused Laboratory Scope

Method validation, analyzer comparisons, reference intervals, quality indicators, preanalytical studies, and instructive clinical case reports.

Credible and Discoverable

All manuscripts publish in English and follow ICMJE and COPE guidelines, with Crossref DOIs and TR Dizin indexing ensuring global discoverability.