Evaluation of Glucose Results Measured by 2 Different Glucometer Brands

1 Department of Biochemistry, Ordu University Faculty of Medicine, Ordu, Türkiye
2 Department of Biochemistry, Memorial Şişli Hospital, Istanbul, Türkiye

Abstract

Objective: The aim of this study was to retrospectively evaluate the performance of 2 glucometer devices using the accuracy criteria specified in ISO 15197:2015.

Methods: Glucose results obtained from Accu-Chek Inform II and Sejoy BG-707 glucometers, which measure according to the glucose dehydrogenase principle, were compared with glucose results obtained from the Cobas 8000 c702 autoanalyzer, which measures using the hexokinase method. Since the minimum number of patients required according to the ISO 15197:2015 protocol is 100, a total of two hundred patients were included using 2 independent samples, with 100 patient results included for each device.

Results: When compared with the autoanalyzer, the Passing-Bablok regression equation for the Accu-Chek brand glucometer was y= 4.606 + .965x, and for the Sejoy brand glucometer y= 6.411 + .964x. In the Bland–Altman plot, the mean difference between glucose values compared to the hexokinase method was for the Accu-Check brand glucometer, 1.41 mg/dL (.08 mmol/L), and for the Sejoy brand glucometer, .09 mg/dL (.005 mmol/L). Furthermore, 98% of the Accu-Chek Inform II results and also 98% of the Sejoy BG-707 results met the ISO 15197:2015 accuracy criteria. In addition, all results for both glucometers were located within Zone A of the Consensus Error Grid.

Conclusion: The results of the study showed that both glucometer measurement results were appropriate and reliable according to the ISO 15197:2015 guideline criteria, demonstrated strong agreement with autoanalyzer results using the hexokinase method, and revealed no statistically significant difference between glucometer and autoanalyzer measurements.

Citation

Noyan, Tevfik, Murat Cihan, Esra Yiğit, and Fatih Ahmet Uzdu. 2026. “Evaluation of Glucose Results Measured by 2 Different Glucometer Brands”. Journal of Turkish Clinical Biochemistry 24 (2):21-26. https://doi.org/10.65717/jtcb.2026.26027.

Publish with JTCB

Diamond Open Access

Fully Diamond OA with no APCs and no submission fees. Your research is freely available to readers worldwide from the day it publishes.

30-Day First Decision

Double-blind peer review from an active editorial team returns a first decision in about 30 days, keeping your manuscript moving.

Global Platform

Backed by the Turkish Society of Clinical Biochemistry, JTCB welcomes submissions from researchers worldwide and reaches an international laboratory audience.

Focused Laboratory Scope

Method validation, analyzer comparisons, reference intervals, quality indicators, preanalytical studies, and instructive clinical case reports.

Credible and Discoverable

All manuscripts publish in English and follow ICMJE and COPE guidelines, with Crossref DOIs and TR Dizin indexing ensuring global discoverability.